The purpose of this research is to check the validation of an auto-analysis method combined with aqua regia digestion apparatus for improvement of water distillation method used as a fluorine test of soil. Fluorine contents of CaO used in the pretreatment course of water distillation method were 120 mg/kg ~ 5,064 mg/kg at the blank test, which was exceeded up to maximum 12.5 times of the soil standard, so it was estimated due to a effect of fluorine existing as impurities of CaO. The recovery test of the same samples indicated that water distillation method and auto-analysis method were 134.5mg/kg and 161.7mg/kg respectively, the recovery ratio of the latter was 16.8% higher than the formal. The validation test of two methods satisfied the standard, but auto analysis method was excellent more than distillation method. Also, auto analysis method could save a analysis time up to maximum 4.7 times by comparison with water distillation method.
Keywords: Fluorine of soil;Water distillation method;Automatic analysis method;Validation;Uncertainty;